Friday, 8 December 2017

Action potential in medical electronics

The Action Potential

Recording an AP requires the isolation of a single cell.
􀂪 Microelectrodes (with tips a few μm across) are used to stimulate and record the
response. A typical AP is 2-4ms long with an amplitude of about 100Mv
The Electroencephalogram EEG
EEG is the graphical representation of the electrical activity of the brain
􀂪 Very commonly used to diagnose certain neurological disorders, such as epilepsy
􀂪 More recently, also investigated whether it can detect various forms of dementia
or schizophrenia
􀃂 EEG is the specific recording obtained using the scalp electrodes from the
surface of the skull
􀂪 During surgery, electrodes may also be placed directly on the cortex. The
resulting signal is then electrocorticogram (ECoG).
􀂪 Just like ECG, EEG is also obtained using several different electrodes places on
different regions of the head / brain
The Event Related
Potentials – ERPs
ERPs are really EEGs obtained under a specific protocol that requires the
patient to response to certain stimuli – hence event related potentials.
􀂪 Also called evoked potentials these signals can be used to diagnose certain
neurological disorders such as dementia, and they can also be used as a liedetector
The oddball paradigm
The guilty knowledge test
Electroretinogram ERG
The ERG is the record of the retinal action currents produced by the retina in
response to a light stimulus.
􀂪 It measures the electrical responses of the light-sensitive cells (such as rods and cones).
The stimuli are often a series of light flashes or rotating patterns
􀂪 The ERG is recorded using contact lens electrode that the subject wears while watching
the stimuli.
Phonocardiogram – PCG
The PCG is the graphic record of the heart sounds and murmurs. It is thus a
mechanical / audio signal, rather than an electrical signal
􀂪 Can be easily heard using a stethoscope
􀂪 Or can be converted into an electrical signal using a transducer
􀂪 Typically used to determine the disorders related to the heart valve, since their
routine opening and closing create the well-known sounds.
• S1 sounds: First heart sounds – ventricular contractions move blood into atria closing
of the AV (mitral and tricuspid) valves, then semilunar valves open and blood ejected
out of ventricles – immediately follows the QRS complex
• S2 sounds: Second heart sounds – Closure of semilunar (aortic and pulmonary) valves
• Any unexpected sound may indicate a malfunctioning valve that causes the blood flow
into / out of a chamber when it should not. Also called heart murmurs.


Define ultrasound

         Mechanical waves in different modalities (longitudinal/lateral) à needs medium to be propagated (solid, liquid, gas)
         > 20 kHz
         Continuous/pulsed
         Spherical/planar/narrow beam/surface wave/Lamb-wave

Physical phenomena behind ultrasound measurements

Transmission
          
         reflection
         transit time
         differences in propagation velocities
         returns to transit time
         doppler-shift in frequency
         flow velocity
         change of acoustic impedance
         comparing to reference
         interference of ultrasound waves (holography)
         interaction of ultrasound and light (photoacousticz)
         ultrasound needs medium for propagation à it doesn’t propagate in vacuum
         because mechanical waves need moving massunits and spring forces between them
         in acoustic emission the medium creates ultrasound (for example, during pressure changes), which is received by sensors
         pulsed mode more common than continuous
         continuous reguires separate transducers for transmitting and receicving
         in pulsed mode an ultrasound burst is sent to the object and the same transducer is switched to listen echoes
         standing wave problem
         in us-therapy pulsed mode gives more effective care without too much heating
The Doppler Equation describes the relationship of the Doppler frequency shift to target velocity.
The frequency difference is equal to the reflected frequency (FR) minus the originating frequency
(FT). If the resulting frequency is higher, then there is a positive Doppler shift and the object is
moving toward the transducer, but if the resulting frequency is lower, there is a negative Doppler
shift and it is moving away from the transducer. In its simplest form it would be calculated as if
the ultrasound was parallel to the target’s direction, as shown in diagram A below.
However, this would be a rare occurrence in clinical practice, because the transducer is rarely
pointed head on to a blood vessel. In real life, the ultrasound waves would approach the target
at an angle, called the Doppler angle ( ). On the following page, diagram B shows the Doppler
equation used in general clinical situations, which includes the Doppler angle.
The Doppler Angle
The ultrasound beam usually approaches the moving target at an angle called the
Doppler angle ( ). This reduces the frequency shift in proportion to the cosine of this
angle. If this angle is known then the flow velocity can be calculated. The equation used
is:
The Doppler Equation
≡ Doppler shift frequency (the difference between the transmitted and
received frequencies)
≡ transmitted frequency
≡ reflected frequency
V ≡ velocity of the blood flow towards the transducer
C ≡ velocity of sound in tissue
θ ≡ the angle between the sound beam and the direction of moving blood
Where:
The Doppler angle ( ) is also known as the angle of insonation. It is estimated by
the sonographer by a process known as angle correction, which involves aligning an
indicator on the duplex image along the longitudinal axis of the vessel.
There are a few considerations that affect the performance of a Doppler
examination that are inherent in the Doppler equation, which are:
– The cosine of 90° is zero, so if the ultrasound beam is perpendicular to the
direction of blood flow, there will be no Doppler shift and it will appear as if
there is no flow in the vessel.
– Appropriate estimation of the angle of insonation, or angle correction, is
essential for the accurate determination of Doppler shift and blood flow velocity.
The angle of insonation should also be less than 60° at all times, since the
cosine function has a steeper curve above this angle, and errors in angle
correction will be magnified.
The simplest Doppler devices use continuous wave (CW Doppler), rather than the pulsed
wave used in more complex devices. CW Doppler uses two transducers (or a dual element
transducer) that transmit and receive ultrasound continuously. The transmit and receive
beams overlap in a Doppler sample volume some distance from the transducer face, as
shown in the diagram below.
volume) is the region of transmitting and receiving beam overlap (shaded region).
Because there is continuous transducer transmission and reception, echoes from all
depths within the area arrive at the transducer simultaneously.
So although CW Doppler can determine the direction of flow, it cannot discriminate
the different depths where the motion originates. The usefulness of CW Doppler
devices is limited, but they are used clinically to confirm blood flow in superficial
vessels, as they are good at detecting low velocities. As they are easily portable, this
can be done at the bedside or in the operating room. Most other clinical applications
require pulsed wave Doppler.

Pulsed Wave Doppler (PW Doppler)https://naveenkumaropm.wixsite.com/technewstamil

Medical Electronics short notes



Half Cell Potential
A characteristic potential difference established by the electrode and its surrounding electrolyte which depends on the metal, concentration of ions in solution and temperature (and some second order factors) .
Half cell potential cannot be measured without a second electrode.
The half cell potential of the standard hydrogen electrode has been arbitrarily set to zero. Other half cell potentials are expressed as a potential difference with this electrode.
Reason for Half Cell Potential : Charge Separation at Interface
Oxidation or reduction reactions at the electrode-electrolyte interface lead to a double-charge layer, similar to that which exists along electrically active biological cell membranes.
Measuring Half Cell Potential


Polarization
If there is a current between the electrode and electrolyte, the observed half cell potential is often altered due to polarization.
Nernst Equation
When two aqueous ionic solutions of different concentration are separated by an ion-selective semi-permeable membrane, an electric potential exists across the membrane.
The Nernst equation for half cell potential is
 






where E0 : Standard Half Cell Potential        E : Half Cell Potential
           a  : Ionic Activity (generally same as concentration)
           n  : Number of valence electrons involved     
Polarizable and Non-Polarizable Electrodes
Perfectly Polarizable Electrodes
These are electrodes in which no actual charge crosses the electrode-electrolyte interface when a current is applied. The current across the interface is a displacement current and the electrode behaves like a capacitor. Example : Ag/AgCl Electrode
Perfectly Non-Polarizable Electrode
These are electrodes where current passes freely across the electrode-electrolyte interface, requiring no energy to make the transition. These electrodes see no
Over potentials. Example : Platinum electrode
Example: Ag-AgCl is used in recording while Pt is use in stimulation

Equivalent Circuit
Cd     : capacitance of electrode-eletrolyte interface
Rd     : resistance of electrode-eletrolyte interface
Rs       : resistance of electrode lead wire
Ecell  : cell potential for electrode
Electrode Skin Interface
Motion Artifact
Why
When the electrode moves with respect to the electrolyte, the distribution of the double layer of charge on polarizable electrode interface changes. This changes the half cell potential temporarily.
What
If a pair of electrodes is in an electrolyte and one moves with respect to the other, a potential difference appears across the electrodes known as the motion artifact. This is a source of noise and interference in biopotential measurements
Motion artifact is minimal for non-polarizable electrodes
Body Surface Recording Electrodes

  1. Metal Plate Electrodes (historic)
  2. Suction Electrodes
(historic interest)
  1. Floating Electrodes
  2. Flexible Electrodes
Commonly Used Biopotential Electrodes
Metal plate electrodes
        Large surface: Ancient, therefore still used, ECG
        Metal disk with stainless steel; platinum or gold coated
        EMG, EEG
        smaller diameters
        motion artifacts
        Disposable foam-pad: Cheap!

(a) Metal-plate electrode used for application to limbs.
(b) Metal-disk electrode applied with surgical tape.
(c)Disposable foam-pad electrodes, often used with ECG
Commonly Used Biopotential Electrodes
Suction electrodes
-           No straps or adhesives required
-           precordial (chest) ECG
-           can only be used for short periods
Floating electrodes
-           metal disk is recessed
-           swimming in the electrolyte gel
-           not in contact with the skin
-           reduces motion artifact


Sunday, 3 December 2017

American tourister bag quality


AMERICAN TOURISTER:

     This is a story behind a bag a normal start-up company decided to produce good quality of bag they started company called American tourister.

     Initially they don't have an idea about the product that they are going to launch tmthey started collecting information on user really needed and how they going to use the product for future purpose so they designed the bag by well finished water resistance bag and they sold on market within few months they got good response from customer then they started doing the business with good quality of product with affordable prize.

     They reached peak in marked then decided to make back bags that came big success in the market they sold more product and now they become no 1 company.

Thank you guys comment if you like support me to do more.